SSB phase noise and the floor problem

Single-sideband phase noise L(f) is given in dBc/Hz versus offset. When measuring directly, the spectrum analyzer reads the power sum of the DUT noise and the analyzer's own noise. The working rule: the analyzer's phase noise at the offset of interest should be 10dB better than the DUT for the reading to be trustworthy; at only 3dB difference the combined reading is already about 1.6dB high. Before choosing a method, put the DUT spec and the analyzer spec into the same table.
Direct spectrum method
Read the noise power around the carrier directly on a spectrum analyzer. It is the simplest setup — one instrument; the hard limit is the analyzer's own floor, and readings fail as they approach it. It suits signals whose phase noise is clearly worse than the analyzer's, and fast relative comparisons.
Two details matter: the resolution filter shape factor (nominal RBW vs equivalent noise bandwidth), and correct use of video averaging — video average lowers the displayed noise, not the true floor.
Comparing published values is the first step of method selection. As an example from JXBY's catalog: signal generators publish typical phase noise of ≤-93 / ≤-98 / ≤-115dBc/Hz@10kHz (three tiers), and the handheld spectrum analyzer publishes -95dBc/Hz@10kHz (typical). Side by side, it is immediately visible which instrument can directly measure which class of signal — and it is the canonical case for separating typical from guaranteed values: acceptance testing should rely on guaranteed values.
Correlation method

The idea: feed the DUT into two measurement channels and cross-correlate. The DUT's noise is correlated between channels; the two analyzers' own noise is not — correlation pushes the analyzer floor down. The improvement grows with the number of correlation segments; common practice yields improvement on the order of 15–25dB, at the cost of a much longer measurement time.
It suits signals at or below a single analyzer's floor: high-stability oscillators, radar local oscillators, low-phase-noise references.
Cross-correlation and dedicated analyzers
Dedicated phase noise analyzers productize cross-correlation: built-in dual channels, automatic reference handling, direct L(f) output. They solve the "repeatable in engineering practice" problem and are the default choice for low-phase-noise work when budget allows. The direct method stays in the toolbox as the cross-check: when both methods agree within their respective error bounds, the result is convincing.
Common error sources

- Resolution filter shape: nominal RBW vs equivalent noise bandwidth directly affects the conversion
- Carrier leakage: mixer LO leakage raises near-offset readings
- AM noise and AM-to-PM conversion: amplitude noise read as phase noise through device nonlinearity
- Vibration and airflow: invisible contributors to close-in phase noise; the isolation platform is not decoration
- Cable phase drift: the reference channel drifts with temperature over long runs
Choosing a method by scenario
- DUT 10dB worse than the analyzer floor → direct spectrum
- DUT at or better than the analyzer floor (high-stability OCXO, radar LO) → correlation / cross-correlation
- Production sorting → direct spectrum + statistical process control
- R&D precision measurement → dedicated analyzer, periodically cross-checked with the direct method
FAQ
- Can a spectrum analyzer measure phase noise directly?
- Yes, but limited by its own floor: readings are trustworthy when the DUT is 10dB worse than the analyzer; near the floor they fail, and correlation methods take over.
- How much can the direct method over-read?
- When the DUT and analyzer floors differ by only 3dB, the summed reading is about 1.6dB high; the smaller the gap, the larger the bias. It is physics, not operator error.
- Do typical vs guaranteed values matter for acceptance?
- Typical values describe average units; guaranteed values are the limit every unit meets. Phase noise varies between units, so acceptance and sourcing comparisons should rely on guaranteed values; typical values serve initial screening.
- Why does correlation take so long?
- Correlation accumulates statistics over long captures: more segments give a lower floor at the cost of time. A 15–25dB improvement typically comes with minutes-to-hours of measurement.