Frequency range: how far up you can measure

The frequency range sets the upper and lower limits. Signals below the lower limit are invisible; signals above the upper limit alias in.
Published frequency ranges of JXBY handheld spectrum analyzers:
| Model | Frequency range | Max real-time BW | Resolution BW |
|---|---|---|---|
| JXBY-080 | 9kHz–20GHz | 40MHz | 1Hz–10MHz |
| JXBY-200 | 9kHz–26.5GHz | 40MHz | 1Hz–10MHz |
| JXBY-260 | 9kHz–1.8GHz | — | 1Hz–3MHz |
Selection rule: set the upper limit by the highest operating frequency of the system under test, then leave one tier of margin below (since real measurements often include harmonics and spurs).
Frequency resolution: how close two signals can be
Frequency resolution determines the ability to distinguish two closely spaced signals.
The corresponding signal-generator parameter is “frequency resolution”, published in two grades: 0.23Hz and 0.01Hz.
Typical consequences of insufficient resolution: missing an adjacent narrowband signal, or misreading two signals as one. If you measure multipath signals, sidebands or narrowband filters, insufficient resolution makes the whole measurement meaningless.
Phase noise: how weak a signal you can dig out

Phase noise is the spectrum analyzer’s noise floor — it determines how much signal you can dig out of how much noise.
Published examples:
- Signal generators: ≤-93dBc/Hz@10kHz, ≤-98dBc/Hz@10kHz, ≤-115dBc/Hz@10kHz (typical)
- Handheld spectrum analyzer: -95dBc/Hz@10kHz
The difference between typical and guaranteed values matters: typical is what a typical sample achieves, guaranteed is the floor all products meet. For strict acceptance, look at the guaranteed value.
Real-time bandwidth: how fast a signal you can see
Real-time bandwidth is the maximum bandwidth the analyzer can capture continuously — it determines how fast a signal you can observe and whether you can see transients like frequency hopping or modulation.
The published max real-time bandwidth of the handheld spectrum analyzer is 40MHz, with resolution bandwidth 1Hz–10MHz.
Resolution bandwidth (RBW) affects frequency resolution; real-time bandwidth affects time resolution — improving one comes at the expense of the other, an inherent trade-off of spectrum measurement.
Tracking generator: turning the analyzer into a frequency-response tester

A tracking generator (TG) is the analyzer’s built-in swept source; with a detector it can measure the frequency-response curves of filters, amplifiers and cables.
The JXBY handheld spectrum analyzer lists “tracking generator: optional” — not standard, must be ordered as an option.
If you will stimulate rather than only observe, the tracking generator is a must; for observation only, it is unnecessary. This item is often missed at order time.
Handheld or benchtop?
Handheld analyzers are portable, start fast and suit the field; benchtops offer dynamic range, display resolution and built-in analysis.
A common workflow: use the handheld in the field to locate the band and interference source, then make precise measurements on the benchtop back in the lab.
Also, the handheld’s displayed average noise level is published as -160dBm — this figure determines weak-signal detection capability.
FAQ
- Which parameters matter most in spectrum analyzer selection?
- Three groups: coverage (frequency range — what you can measure), resolving power (frequency resolution and phase noise — how close signals can be separated and how weak signals can be found), and real-time capability (real-time bandwidth and RBW — how fast a signal you can see). Also confirm whether a tracking generator is needed.
- What goes wrong when frequency resolution is insufficient?
- You miss adjacent narrowband signals, or misread two close signals as one. When measuring multipath signals, sidebands or narrowband filters, insufficient resolution makes the whole measurement meaningless. JXBY signal generators publish frequency resolution grades of 0.23Hz and 0.01Hz.
- What is the difference between typical and guaranteed phase noise?
- Typical is what a typical sample achieves; guaranteed is the floor all products meet. Strict acceptance looks at the guaranteed value. For example, the published handheld figure is -95dBc/Hz@10kHz — confirm whether that is typical or guaranteed when comparing.
- Is a tracking generator required for swept measurements?
- Yes. The tracking generator is the analyzer’s built-in swept source; with a detector it measures frequency response of filters, amplifiers and cables. For observation only it is unnecessary. Note the JXBY handheld’s tracking generator is optional — state it explicitly when ordering.
This guide is based on the official product pages and solution descriptions /products/spectrum-analyzer/. Parameters are quoted verbatim, without secondary processing.
We provide pre-sales selection consulting, on-site solution planning, technical support, operator training and after-sales calibration services. Phone / WeChat / WhatsApp: +86-15626514602 Email: sales@jxby.tech Service hours 08:00–22:00.
Address: Room 412, Building 1, Phase 1, Dongjiu Innovation Technology Park, No. 76 Bulan Road, Xialilang Community, Nanwan Street, Longgang District, Shenzhen, Guangdong, China