Test range up to 100 MHz-500 GHz for reflectivity and shielding effectiveness, built around a vector network analyser with an arc-shaped or adjustable-radius bow frame, test cables, antennas and system software.

The material absorption and shielding performance test system covers 100 MHz to 500 GHz and measures material reflectivity and shielding effectiveness. For absorption testing the system comprises a vector network analyser or microwave integrated tester, a semi-circular or radius-adjustable (L-type) bow frame, test cables, test antennas, other microwave components and system software. Depending on the required frequency and material performance, reflectivity is measured with the bow-frame method or the portable reflectivity method. For shielding effectiveness the system comprises a flanged coaxial fixture or a shielded room, test cables, test antennas, other microwave components and system software, with the flanged coaxial method or the shielded-room method selected to suit the requirement.
Bow-frame method · Portable reflectivity method · Shielding effectiveness method · System software
For absorption-material reflectivity testing, calibration and measurement can be carried out at different incidence angles. Each axis is controlled independently, supporting fixed-angle incidence and extraction of scattered signals at different angles.
Fast in-situ reflectivity measurement of absorption materials from 2 GHz to 18 GHz. Operation is simple and repeatability is good, with accuracy comparable to the far-field bow-frame method. The system is ultra-wideband and compact, supports on-site testing (a tablet can replace the portable PC, mounted on a custom antenna bracket), and suits coating-type absorption materials.
Suited to fast shielding-effectiveness testing of sheet samples (flanged coaxial method) and block samples (shielded-room method) in the relevant band. The shielded-room method offers a wider test band and greater dynamic range.
The software provides the bow-frame and flanged coaxial methods with a rich instrument resource library and compatibility with mainstream instruments at home and abroad. The setup interface is simple and friendly, and measurement data can be post-processed. Functions include test-method selection with the corresponding computation, reflectivity and shielding calculation from the two-port reflection and transmission coefficients of the sample, electromagnetic parameter display in a two-dimensional coordinate system, and result analysis that loads multiple measurement files and overlays several curves. Data from the same band can be compared across repeated measurements to judge repeatability; data from different bands can be plotted together to show how the electromagnetic parameters vary across a wide band.
The material reflectivity and shielding performance test system offers fast, wideband measurement for research institutes and manufacturers developing and producing absorbing and shielding materials. Typical test objects are plate materials, structural materials, woven materials and other composites.
Flexible configuration, complete test methods
Bow-frame calibration & test at different incidence angles
Portable in-situ reflectivity measurement
Shielded-room method: wide band, high dynamic range
Easy software, compatible with multiple fixtures
Send us the device under test, frequency range, sample size and site conditions. Our engineering team will provide an executable configuration during solution consulting and site planning.